Fault judgment and solution of wireless hanging scale signal sometimes without

Judgment and Solution to the Failure of Wireless Hanging Scale Signal

        The signal of the wireless hanging scale is sometimes unavailable, especially for the old scale that has been used for several years. First, check the antenna. The elastic copper wire in the antenna is welded to the base. Sometimes the antenna looks like an antenna, but it has lost its function. Then, check the antenna base carefully to see if there is any oil stain, dust, or defective replacement.

        Moreover, check the contact between the battery plug and the scale body socket. After the plug is inserted, gently rotate the plug to see if the finger light will turn on and off for a while to judge whether the contact is good. If the contact is not good, the power will be cut off and powered on for a while during the use. Because the scale is often in a dynamic state when it is used, there will be a signal for a while and there will be no signal for a while. For temporary solutions, you can take a small screwdriver and gently pry the central axis in the socket, Make it slightly biased, increase the contact strength with the socket, and it is better to replace it for a long time.

        If all the above are normal, and the signal of the suspension scale is still absent from time to time, it can only be a problem of ad transmission, and only a replacement test is needed.

        Generally, for signal problems that cannot be determined the exact cause, it can be divided into two steps. First, replace the antenna, battery cable and socket, and then use them. It can basically solve 80% of these problems, and the only other problem is to replace the AD transmitter box.

        Generally, if the quality of the instrument is OK after a long journey back to the factory for maintenance, it is recommended to replace the inside completely, leaving only the shell and sensor, so as to avoid twice of maintenance and delay in use. Of course, the cost of replacement is also high, so it is suitable for old scales that have been used for several years.

        The above troubleshooting only fails to mention the problem of the display. Up to now, it is rare for the instrument to have no signal except for the antenna. It may be that there is no vibration on the instrument or the user's use and maintenance are good, so there is basically no fault.



Shanghai Quanfu Industrial Co., Ltd. can provide solutions to meet all your weighing needs. We look forward to working with you!


Prev:Application and selection of counting scale
Next:Electronic balance selection, not only accuracy and range

Product Categories

QQ

Mobile
Consult